ASTM Mock Server Access Guide

Quick Access

The ASTM mock server is accessible via multiple methods:

From Host Machine

# Direct connection
nc localhost 5000

# Or via Python
python -c "import socket; s=socket.socket(); s.connect(('localhost', 5000)); s.send(b'\x05'); print(s.recv(1).hex())"

From OpenELIS Backend Container

  • Container name: openelis-astm-simulator:5000
  • IP Address: Check with docker inspect openelis-astm-simulator --format '{{range .NetworkSettings.Networks}}{{.IPAddress}}{{end}}'

Via Nginx TCP Proxy

  • Port: 5000 (proxied through nginx)
  • Access: localhost:5000 or analyzers.openelis-global.org:5000 (if DNS configured)

Test Analyzer Configuration

A test analyzer (ID: 1000) is pre-configured to connect to the ASTM server:

SELECT a.id, a.name, ac.ip_address, ac.port, ac.status
FROM analyzer a
JOIN analyzer_configuration ac ON a.id = ac.analyzer_id
WHERE a.id = 1000;

Current Configuration:

  • Name: Hematology Analyzer 1
  • IP: 172.20.1.6 (container IP - auto-detected)
  • Port: 5000
  • Status: SETUP

Updating Analyzer Configuration

If the ASTM server container IP changes, update the analyzer:

# Get current IP
ASTM_IP=$(docker inspect openelis-astm-simulator --format '{{range .NetworkSettings.Networks}}{{.IPAddress}}{{end}}')

# Update analyzer configuration
docker exec openelisglobal-database psql -U clinlims -d clinlims -c \
  "UPDATE analyzer_configuration SET ip_address = '$ASTM_IP', port = 5000 WHERE analyzer_id = 1000;"

Nginx Stream Configuration

The nginx proxy includes a TCP stream proxy for the ASTM server:

stream {
    server {
        listen 5000;
        proxy_pass openelis-astm-simulator:5000;
        proxy_timeout 30s;
    }
}

This allows the ASTM server to be accessed through the nginx proxy on port 5000.

Testing Connection

Test ENQ/ACK Handshake

import socket
s = socket.socket()
s.connect(('localhost', 5000))
s.send(b'\x05')  # ENQ
response = s.recv(1)
assert response == b'\x06'  # ACK
print("✓ Connection successful!")
s.close()

Test from OpenELIS UI

  1. Navigate to /analyzers
  2. Select "Hematology Analyzer 1" (ID: 1000)
  3. Click "Test Connection"
  4. Should see "Connection successful" within 30 seconds

Troubleshooting

Server Not Responding

# Check container status
docker ps | grep astm-simulator

# Check logs
docker logs openelis-astm-simulator

# Test direct connection
docker exec openelis-astm-simulator python -c "import socket; s=socket.socket(); s.bind(('0.0.0.0', 5000)); print('Port 5000 is listening')"

Nginx Proxy Not Working

# Check nginx config
docker exec openelisglobal-proxy nginx -t

# Check stream module
docker exec openelisglobal-proxy nginx -V 2>&1 | grep stream

# Restart proxy
docker restart openelisglobal-proxy

Analyzer Connection Fails

  • Verify analyzer IP matches container IP: docker inspect openelis-astm-simulator --format '{{range .NetworkSettings.Networks}}{{.IPAddress}}{{end}}'
  • Check analyzer configuration in database
  • Verify ASTM server is running: docker ps | grep astm