Test of v1.5.11 with additional udc default state changes from
- Add devices required for Hyperion beamstop check dodal#1695
- Check beamstop using ipin diode on udc_default_state #1429
Testing schedule
Default state
- Review by beamline staff of default state sequence before testing DONE
- Check cryostream in HC1 causes default state to be rejected PASS
- Check that sample present causes beamstop check to fail
- Check hutch shutter is opened by default state plan
- Check that scintillator is moved out (can only do this if ap-sg is parked)
- Check that fluorescence detector moved out PASS
- Collimation table is homed PASS
- All of above can complete while beamstop check?
- Beamstop check
- Sample shutter closed PASS
- backlight out PASS
- detector to within range PASS
- feedback on, transmission to 100% - commissioning mode
- ap-sg to out-of-beam PASS
- beamstop to out-of-beam PASS
- sample current xbpm1 checked at least minimums (may need to tie this to commissioning pv)
- ipin gain to 10e4 low noise PASS
- detector shutter closed PASS
- all of above to complete except detector motion PASS
- background ipin reading - test fails if below threshold - PASS
- beamstop to data collection PASS
- double check sample shutter and detector shutter closed
- wait for detector motion to complete
- sample shutter open
- 2nd ipin reading
- check delta-ipin does not exceed threshold
- On successful beam check:
- Select small aperture
- coarse + fine cryostream in
Baton release
- Detector shutter is closed when baton is released