Describe the bug
In the event of a failure of the nut trap, there is no 'backup' mode for removing the device, resulting in a situation where the device has to be destroyed while still on the patient to be removed.
Expected behavior
Even when the nut trap fails, there should be a graceful way to remove the nut.
Failure Mode
The nut cannot be removed when the nut trap is stripped or fails.
Effect
When the nut cannot be removed, the device is stuck on the patient and more drastic efforts have to be taken to remove the device.
Cause
The nut trap completely encloses the nut.
Mitigation / Control
Cut the nut trap height in half to expose some of the nut.