|
6 | 6 | % the Hyper spectral line-scanning microscope (HSM) by using digital signals. This utilizes a |
7 | 7 | % National Instruments (NI) data acquisition (DAQ) device to send a 16-bit digital signal to |
8 | 8 | % the galvo controller. The galvo controller then converts the 16-bit digital signal to an analog voltage that adjusts |
9 | | - % the galvo mirror's angle (Range:[-15, 15]) for scanning purposes. It |
10 | | - % changes the angle of the galvo mirror to scan the sample. The position |
11 | | - % of the galvo mirror is determined by a 16-bit digital signal |
12 | | - % (Word property). This signal can represent integer values ranging |
13 | | - % from 0 to 65535, corresponding to the full range of movement of |
14 | | - % the mirror. The galvo mirror is driven by 16 digital channels |
15 | | - % configured on the NI DAQ device. These channels send the digital |
16 | | - % word to control the mirror's position. |
| 9 | + % the galvo mirror's angle (Range:[-15, 15]) from the center position with a linear response to analog voltage from |
| 10 | + % -10 to 10 Volts. The NI-DAQ card (PCIe-6343) sends a 16-bit binary signal to the control board to set the mirror angle. |
| 11 | + % This galvo mirror can be also controlled with an analogy servo driver (e.g. Cambridge Technology 673XX). |
| 12 | + % However, the NI-DAQ card we used can only output an analog voltage with an Absolute Accuracy at Full Scale of 3 mV, |
| 13 | + % which is larger than the required scanning step of 1 mV for our application. Therefore, we selected the digital servo |
| 14 | + % driver for this application. Furthermore, the scanning step size is determined by the magnification of the imaging system, |
| 15 | + % from our optical design, moving the beam by 100 nm at the sample plane requires galvo to rotate by 0.00184 degree, |
| 16 | + % which corresponds to applying 1mV if using an analog driver. |
17 | 17 | % |
18 | 18 | % ## Class Properties |
19 | 19 | % |
|
0 commit comments