Description
Overview
When attempting:
cold_module_pbs = protocol.load_module("temperature module gen2", "C1")
cold_mod_pbs_adapter = cold_module_pbs.load_adapter("opentrons_96_deep_well_temp_mod_adapter")
cold_reservoir = cold_mod_pbs_adapter.load_labware(
"agilent_1_reservoir_290ml"
)
the resulting error happens:
ProtocolCommandFailedError [line 80]: Error 4000 GENERAL_ERROR (ProtocolCommandFailedError): LabwareCannotBeStackedError: Labware agilent_1_reservoir_290ml cannot be loaded onto labware opentrons_96_deep_well_temp_mod_adapter
This is just incorrect, because this labware fits quite nicely onto your opentrons_96_deep_well_temp_mod_adapter (I've done it IRL!). I know I can just make my own custom labware, but I think it'd be helpful to update the .json on the labware library "agilent_1_reservoir_290ml "
Steps to reproduce
cold_module_pbs = protocol.load_module("temperature module gen2", "C1")
cold_mod_pbs_adapter = cold_module_pbs.load_adapter("opentrons_96_deep_well_temp_mod_adapter")
cold_reservoir = cold_mod_pbs_adapter.load_labware(
"agilent_1_reservoir_290ml"
)
Current behavior
ProtocolCommandFailedError [line 80]: Error 4000 GENERAL_ERROR (ProtocolCommandFailedError): LabwareCannotBeStackedError: Labware agilent_1_reservoir_290ml cannot be loaded onto labware opentrons_96_deep_well_temp_mod_adapter
Expected behavior
Should be fine!
Operating system
Windows
System and robot setup or anything else?
OT Flex v8.2