Skip to content

Agilent_7900_ICPMS #58

@kurtw555

Description

@kurtw555
<style> </style>

Parser Name: Agilent_7900_ICPMS

Example Data File: 20200303 SPLP TCLP CMT ICP-MS Batch Report

Instrument File Field Location Universal Template File Field Location Special Notes
Column G (Beginning with row 3) Column A (Aliquot)  
Column D (Beginning with row 3) Column F (Analysis Date/Time)  
Row 1 (Beginning with column H) Column B (Analyte Identifier) The number of columns may vary. Column H through the last column (Column EM in the example data file I have attached) will need to be parsed
Column H (Beginning with row 4) Column C (Measured Value) Convert blank cells, “N/A”, and “<0.00” to be imported as “0”   Beginning with Column H of the instrument file, every other column will need to be parsed. For example, Column H values will need to be mapped starting in row 3, Column I will be skipped, Column J will be mapped, and Column K will be skipped, etc. If it is any easier for you, cell 2 in the column that needs parsed will read either “Conc. [ ppb ]” or “CPS” whereas cell 2 in the column that will be skipped will read either “Conc. RSD” or “CPS RSD”   The number of columns may vary. Column H through the last column (Column EM in the example data file I have attached) will need to be parsed

Metadata

Metadata

Assignees

No one assigned

    Labels

    Type

    No type

    Projects

    No projects

    Milestone

    No milestone

    Relationships

    None yet

    Development

    No branches or pull requests

    Issue actions