Phase 10 — ECG image digitization: fidelity report

Round-trip reconstruction fidelity over 100 random PTB-XL records, measured by scripts/eval_digitization.py: each signal is rendered to a paper-ECG image, digitized back to a (12, T) signal, and compared per lead.

Metric: mean per-lead Pearson correlation between the original and the digitized signal (resampled to a common length + best small-lag alignment, so a global time/gain calibration offset isn't scored as lost shape), plus normalized RMSE (RMSE / original std). corr p10 is the 10th-percentile record — the typical worst case, not the average.

Results

photo level corr (mean) corr (std) corr p10 nRMSE (mean)
0.0 — clean render 0.8848 0.0336 0.8278 0.5101
0.5 — mild phone-photo 0.8151 0.049 0.7459 0.6144
1.0 — heavy phone-photo 0.7056 0.0763 0.6257 0.802

Examples

Rendered paper ECG (the digitizer's input): render

Same, photographed (blur + noise + JPEG): photo

Original vs digitized, three leads: overlay

How it works

Classical computer vision (no training data — no dataset of real paper-ECG photos paired with signals exists), in src/digitization/:

  1. grid — the pink grid's bounding box gives the plotting rectangle; the median spacing between detected grid lines gives the px-per-mm calibration.
  2. trace — an adaptive luminance threshold isolates the dark trace ink from the lighter grid; each of the 12 stacked lead bands is read column-by-column as the darkness-weighted centroid row.
  3. calibrate — pixels -> mm via the grid pitch, mm -> mV / seconds via the standard 10 mm/mV and 25 mm/s; the per-lead baseline is removed and each lead is resampled to the target rate.

Honest limitations

  • Validated on rendered images, not real-world phone photos. The photo levels approximate blur/noise/JPEG but not perspective skew, folds, or shadows; a learned trace/grid segmentation model (trainable on this renderer's paired output) is the natural upgrade for those, and the reason to keep the renderer around.
  • Sharp QRS peaks are the main fidelity loss — raster digitization smooths narrow, near-vertical strokes, so clean fidelity sits around 0.9 rather than ~1.0 (a clean sine round-trips at 0.998).
  • The standard clinical 3x4 mosaic carries only 2.5 s per lead, so full 12-lead x 10 s reconstruction uses the stacked full-width layout; the mosaic is rendered for display only.