Skip to content

BBD302 XY stage axes appear as separate devices #9

Description

@jacob-chamoun

As of May 2025, Thorlabs MLS203-series linear motor driven XY microscope stages are driven by the BBD302 controller. When following the installation and configuration instructions, the X and Y axes of the stage (controller channels 1 and 2) appear in the installed device list as two separate linear stages.

Image

This causes micro-manager to interpret the XY stage as two separate linear stages rather than a single XY stage. Later in the hardware configuration wizard, no option for a default XY stage appears, while the separate X and Y axes incorrectly appear as choices for a default focus stage.

Image

After hardware configuration when running micro-manager, the stages appear in the stage control window as two separate linear stages.

Image

This status breaks the micro-manager multi-dimensional acquisitions and the create grid feature in the stage positions list, which appear greyed-out.

Image

Image

== System ==

Hardware:
BBD302 brushless servo controller
MLS203 High-Speed, Low-Profile Motorized XY Scanning Stage

Software:
Windows 11 Pro
MM Studio version: 2.0.3 20250506
MMCore version 11.5.0
Device API version 73, Module API version 10