Skip to content

Centroid or Profile selection for placed scans #39

@chhh

Description

@chhh

Hi, when I connect to Exploris 480 it lists the following available scan parameters - I don't see anything for centroid/profile selection, is this right?

Name[FirstMass], Selection[40.0-8000.0], DefaultValue[120.0], Help[First m/z, lower edge of displayed spectral range]
Name[LastMass], Selection[40.0-8000.0], DefaultValue[2000.0], Help[Last m/z, upper edge of displayed spectral range]
Name[IsolationRangeLow], Selection[40.0-2500.0], DefaultValue[120.0], Help[Lower m/z edge of isolation range]
Name[IsolationRangeHigh], Selection[40.0-2500.0], DefaultValue[2000.0], Help[Upper m/z edge of isolation range]
Name[MsxInjectRanges], Selection[string], DefaultValue[[]], Help[MSX m/z-ranges for isolation steps. If a single range is exceeding isolation min/max, no range is applied! Format - example:  [(191.5,194.5), (523.3,525.3), (1117.0,1127.0)]]
Name[MsxInjectTargets], Selection[string], DefaultValue[[]], Help[MSX AGC targets for injection steps. If the total sum of 10e6 is exceeded, no definition is applied! Format - example: [20000, 50000, 35000]]
Name[MsxInjectMaxITs], Selection[string], DefaultValue[[]], Help[MSX (max) inject time in milliseconds for isolation steps. If the sum of ITs is exceeding 1000ms, no defition is applied! Format - example:  [80.0, 8.5, 120.0]]
Name[MsxInjectNCEs], Selection[string], DefaultValue[[]], Help[MSX normalized collision energy for isolation steps. Format - example:  [30.0, 55.5, 17.8]]
Name[MsxInjectDirectCEs], Selection[string], DefaultValue[[]], Help[MSX HCD collision energy direct eV setting. Format - example: [10.4, 22.1, 42.8]]
Name[MaxIT], Selection[0.03-1000.00], DefaultValue[10.0], Help[Injection time in milliseconds, maximum injection time for AGC operation. If set to <=0, automatically uses the detect time of the previous scan as maximum inject time.]
Name[Resolution], Selection[7500.0-480000.0], DefaultValue[30000.0], Help[Resolution setting, determines transient data acquisition duration per scan]
Name[Polarity], Selection[0,1], DefaultValue[0], Help[Ion polarity setting, 0=positive ions, 1=negative ions]
Name[NCE], Selection[-200.0-200.0], DefaultValue[0], Help[HCD Collision Energy in normalized units.]
Name[NCE_NormCharge], Selection[1-5], DefaultValue[2], Help[ControlNode]
Name[NCE_Factors], Selection[string], DefaultValue[], Help[Factors for stepped CE for a more flexible definition. Example:  [0.01, 0.5, 0.75, 1.0] (reference is NCE/DirectCE, MAX = 1.0, best select the highest value for NCE/CE)]
Name[DirectCE], Selection[-200.0-200.0], DefaultValue[0.0], Help[HCD collision energy direct eV setting (NCE needs to be set)]
Name[SourceCID], Selection[-135.0-135.0], DefaultValue[0.0], Help[In-source fragmentation energy (eV)]
Name[Microscans], Selection[1-10], DefaultValue[1], Help[Number of microscans per logicel scan.]
Name[AGC_Target], Selection[1-10000000], DefaultValue[1000000], Help[Target number of trapped ions for automatic control (AGC). Active when mode is 'Absolute AGC Target']
Name[AGC_Mode], Selection[0,1], DefaultValue[1], Help[1=Enable automatic injection control for number of trapped ions (AGC), 0=Use fixed injection time]
Name[AGC_PeriodicPS], Selection[0.0-5.0], DefaultValue[0.66], Help[Setting this !=0 inserts wide-mass-range-AGC prescans at given period [sec] (will be active only if prescans are required by analytical scans). The period is calculated automatically when this is set to -1.]
Name[PrOSA_Support], Selection[0,1], DefaultValue[1], Help[1=Enable/0=disable PrOSA periodic operation (for AGC improvement)]
Name[Scan_Handshake_Mode], Selection[0,1,2], DefaultValue[0], Help[0=Disable(default)/1=Enable-for-all/2=Enable-for-MS1-only Peripheral connector B per-scan-handshake functionality]
Name[Scan_Handshake_MinTriggerDelay], Selection[0.0-500.0], DefaultValue[10.0], Help[Handshake min. Trigger Delay (msec). Default is 10.0 msec]
Name[Scan_Handshake_Nth_Counter], Selection[1-1000000], DefaultValue[1], Help[Setting this >1 will insert a handshake only every N-th scan]
Name[Scan_Handshake_Nth_Cnt_Cycle_Terminate], Selection[0-10000000], DefaultValue[0], Help[Setting this to a NEW value >0 will terminate and restart an active N-th-scan-conter-cycle with the next scan. Thus next scan will be handshaked even if N-th counter is not reached.]

Metadata

Metadata

Assignees

No one assigned

    Labels

    No labels
    No labels

    Type

    No type

    Projects

    No projects

    Milestone

    No milestone

    Relationships

    None yet

    Development

    No branches or pull requests

    Issue actions