You must calibrate the volume of the reward given to the mouse. 💧
We typically use 5 microliters per reward, but this task/experiment dependent.
To calibrate your reward:
- Set a starting valve opening duration (in ms), typically 50ms
- Place a 25mL Falcon tube below your lick spout to collect rewards (make sure it's right below)
- Set GUI parameters to:
-
- Association ON in order to give rewards at each trials
-
- Stimulus trials only (anyone)
-
- Trial duration, ITI to 1-2 secs minimum to allow for proper settling of water in the tube (that way you calibration settings are more similar to taht during an actual behavioural session)
- Run GUI for e.g. 1000 trials
- Reward volume per trial (uL) = (total volume collected * 1000) / number of trials
- If this is not the target reward volume, repeat adjusting the valve opening time