Rigol DM3058 Calibration Project

The Rigol DM3058E digital multimeter's Calibration/Alignment procedure is undocumented. This repo is to support research into automating both Calibration testing and creating the possibility of Alignment without having to send equipment to Rigol.

DISCLAIMER

Alignment should only be preformed if you have the equipment and skills required! If you don't know what you are doing you will end up with a LESS accurate meter, not a more accurate meter

Definitions

  • Calibration - check if system is within spec, do not modify the system.
  • Alignment - modify the system so it is back with spec.

State of the project

Currently a partial picture of the alignment process has been uncovered. After partially analyzing the firmware, the following commands can be issued through the SCPI USB interface.

:CALI:STEP <n>
:CALI:STATUS?
:CALI:<MODE>:ZERO <Range>
:CALI:<MODE>:GAIN <Range>

Where

  • n is a number between 0 and 492 (with some gaps)
  • mode is anything you could put after a :MEAS scpi commands (eg VOLT:DC)
  • Range 0,1,2,... corresponding to the Range being calibrated

Commands in the style :CALI:VOLT:DC:ZERO 0 work great, but can not currently be stored to non volatile memory. they cant survive *RST

commands in the style :CALI:STEP n can be stored via :CALI:STEP 487 and be reset to factory defaults via :CALI:STEP 488 but it is not as simple as zero and gain there is some extra piece that seams to add in a non linearity and requires more research.

Steps >487 are important for writing and resetting values 487. seems to write the values into flash memory 488. resets the values to a built in default. 489. not clear but references AC half Ranges 490. prints the calibration constants (must use a raw read) 491. seems to disconnect the front terminals (hear a relay click) 492. seems to reconnect the front terminals (hear a relay click)

single elements of calibration can be accomplished but the entire procedure has not been mapped out yet.

before running calibration code the first few elements of the calibration constants were

In [47]: printCalibration(dumpCalibration())
b'00000000'
       0.00        1.00        1.00        1.00        1.00        0.00
# this is intentionally aligning this to incorrect values to prove a point.
# to undo this  run step 488
dmm.range=0
cal.setVoltage(.01) #should really be a short (see above)
dmm.calibrate('step 0') # cal ZERO
cal.setVoltage(.1) # should really be 200mV
dmm.calibrate('step 34') # cal GAIN
dmm.calibrate('step 487') # write Vals

this results in:

In [45]: printCalibration(dumpCalibration())
b'00000000'
      -0.02        3.09        3.09        1.00        1.00        0.00

I'd call that a successful test. but it doesn't really work as well as it sound (introduces serious nonlinearity in the readings.)

Side benefits

A few commands were discovered that are not directly related to calibration but are interesting none the less.

":WEBcontrol:KEY:SET <n>" //tries to press a key but keys cant be pressed while remote is enabled
":WEBcontrol:KEY:GET?" // returns a bitmask related to keys that are currently lit. see the Expiriments page for mask values.
":WEBcontrol:GUI:GET?" // returns a bitmap of the current screen.
"SYSTEM:PROD:PASS rigolproduct" // allow you to change things you really shouldent
"SYST:PROD:SETType DM1234" // change model type (DM3058E,DM3068)
"SYSTEM:PROD:SETSerial <newserial>" // change your serial number (ie DM3R1234A1234)
"SYST:PROD:SETMacaddr aa:AA:aa:aa:aa:aa" // only really matters if you have an LXI interface...

"SYSTEM:PROD:CLOSE rigollan" // save those things you really shouldent change

"WINDOW:STATUS 0|1|ON|OFF" // turn off entire display
"WINDOW:STATUS?" // check if display is on

To save a png of the screen use SaveScreen("output.png")

Cali:Step

the steps seem to be split into groups based on mode and ranges

internally modes are numbered: 0. Volt DC (5 ranges)

  1. Volt AC (5 ranges)
  2. Amps DC (6 ranges)
  3. Amps AC (4 ranges)
  4. Resistance (7 ranges)
  5. 4 Wire Res (7 Ranges)
  6. Frequency
  7. Period
  8. Continuity
  9. Diode

The steps seep to follow that the setting byte is related to if step is a Zero or gain step but toward the bottom of the list there is obviously more going on. (1 seems to be ZERO, 0 seems to be GAIN).

most steps setup some internal memory locations then let the multimeter chug along.

//bad pseudocode
case n:
  IsCalRunning=1
  removeCalConstants??=1
  ZeroGain=<settingByte>
  SCPIprintf("CALIBRATION:STEP %d",n)
  break:
Start Stop Length Setting Byte Note
0 4 5 1 VDC Zero (Ranges: 200mV, 2V, 20V, 200V, 1kV)
10 14 5 1
19 25 7 1 Res2? SHORT
26 32 7 1 Res4? SHORT
33 33 1 1 Frequency Calibration (give 100kHz @ 2V) (DC will hang the dmm)
34 38 5 0 VDC gain
105 109 5 0
110 114 5 0
116 119 4 0
120 126 7 0 Res2? gain?
127 133 7 0 Res4? gain?
140 140 1 0 hangs without VAC input (tweaks Freq measurement)
141 141 1 0 hangs without VAC input (tweaks Freq measurement)
142 146 5 10 vdc related?
147 151 5 10
152 158 7 10 Res2?
159 165 7 10 Res4?
361 361 1 11
362 366 5 12 effects CalConstant[3]
367 371 5 13 effects CalConstant[3&4]
372 376 5 12 effects CalConstant[3]
377 381 5 13 effects CalConstant[3&4]
382 386 5 14 AC and DC HANG
387 391 5 15 NO DC (hangs)
392 396 5 14
397 401 5 15
402 408 7 14 Res2?
409 415 7 14 Res4?
416 416 1 1
417 417 1 1
418 418 1 10
419 419 1 13
420 420 1 12
421 421 1 14
422 422 1 15
423 427 5 16
428 432 5 17
433 437 5 19
438 442 5 20
443 447 5 21
448 448 1 18
449 452 4 16
453 456 4 17
457 460 4 19
461 464 4 20
465 468 4 21
469 474 6 1 CAPacitance Zero?
475 480 6 10 CAPacitance Gain?
481 486 6 0
487 487 1 NA Write Calibrations to factory memory
488 488 1 NA Reload default constants
489 489 1 NA Half Range related?
490 490 1 NA Dump Calibration Constants
491 491 1 NA DMA4 coms(Disable Connection to terminal?)
492 492 1 NA DMA4 coms(Enable Connection to terminals?)

The last 5 steps are very interesting. and have some impressive effects.

internal function representations

functions are set in multiple places and do _setfunction

  • VDC 1

  • VAC 2

  • ADC 3

  • AAC 4

  • ReS2 5

  • res4 9

  • freq A

  • peri 6

  • cont 7

  • diode b

  • cap ???

  • setRange - sets dmmMode

  • VDC 0 - 0x1100

  • VAC 1 - 0x3100

  • ADC 2 - 0x2100

  • AAC 3 - 0x4100

  • res2 4 - 0x5100

  • res4 5 - 0x6100

  • cont 6 - 0x9100

  • diode 7 - 0xA100

  • freq 9 - 0x7100

  • Period a - 0x8100

  • cap b - 0xe100

Repo code

helper classes were created to make my life simpler while testing out theories.

  • Fluke5500 is a calibrator that can generate the signals for the DM3058E
  • this, or similar 5520A, should be used if you want to actually align your multimeter
  • DP832 is a multi channel power supply. This is not a precision instrument don't calibrate your DMM with this!
  • it is smaller than the Fluke5500 and fits on my desk, it is enough to figure out if i have actually realigned the DMM
  • DM3058E is the multimeter under investigation.
  • DM3058.py holds scripts that test alignment and calibration.

Setup (Windows)

Generally a project like this would use Linux. I don't have a dedicated Linux box and NIVISA doesn't play nice with Windows Subsytem for Linux (WSL). This is developed and aimed at Windows with python3

Install nivisa

https://pyvisa.readthedocs.io/en/latest/faq/getting_nivisa.html#faq-getting-nivisa

Clone the Repo

git clone https://github.com/bveina/DM3058E-Calibration.git
cd DM3058E-Calibration

Setup your virtual env (this is the windows version of virtual envirment)

virtualenv venv
venv\Scripts\activate
pip install -r requirements.txt

Enter the terminal

ipython -i testbed.py

This will attempt to connect to the fluke and the DMM. you may need to adjust your code to speak to the correct device.

About

Scripts to work with the DM3058 over the USB-SCPI interface